Selection Guide for Spectrum Dialysis Membranes
Selecting "the Right Membrane"
The key to a successful dialysis separation is using the RIGHT MEMBRANE under the right conditions. There are 6 steps to follow when selecting a membrane:
Select grade (Biotech or Standard) based on application.
Select type based on chemistry: CE or RC.
Select membrane configuration based on application (tubing, disposable device or flat-sheet system).
Select from among 18 MWCO's based on solute size.
Select from among 5 sizes (disposable device) or from among 16 flat widths (tubing) based on volume.
Select the membrane treatment preferred for your application.
Use Spectrum's Dialysis Parameter Search to help select a product:
Spectra/Por Biotech Grade Membranes
Biotech membranes are specifically engineered for more critical separation constraints in which an ultrapure membrane with a precise MWCO is required. Since these synthetic membranes are manufactured using a process without metal salts, no pre-treatment or cleaning is required. Biotech membranes are available in 2 membrane types:
Biotech Cellulose Ester (CE) wide range of selectivity & purity
Biotech Regenerated Cellulose (RC) combined selectivity, purity & resistance
For ultimate convenience and ease-of-use, Spectrum offers:
Biotech Ready-to-Use Dialysis Devices small volumes, in-process sampling & rapid dialysis
Spectra/Por Standard Grade Membranes
Ideal for basic, multi-purpose dialysis, this low cost membrane can be used for a wide variety of dialysis applications in which only general separations are required and low level membrane impurities are of little concern. The standard RC type is composed of natural cellulose regenerated from cotton linters for strength and durability.
Standard Regenerated Cellulose (RC) broad range, low cost lab applications
Membrane Grade: Biotech Standard
Membrane Type: CE RC RC
MWCO (Daltons): 100-500, 500-1000, 3.5-5K, 8-10K, 20K, 50K, 100K, 300K, 1000K 3.5k, 8k, 10k, 15k, 25k 1k, 2k, 3.5k, 5k, 6-8k, 10k, 12-14k, 15k, 25k, 50k
Physical Properties: Opaque, Rigid, Hydrophilic Opaque, Flexible, Hydrophilic Clear Flexible, Hydrophilic
Organic Resistance: Fair Good Good
pH Range: 2 - 9 2 - 12 2 - 12
Temperature Limit: 37°C 60°C 121°C
(Autoclavable)
Flat Sheet: NO NO YES
Recommended
Closures: Universal Only Universal Only Universal or Spectra/Por®
Ready-to-Use
Dialysis Devices: Float-A-Lyzer G2
Tube-A-Lyzer NO Dialysis Sacks
相關(guān)產(chǎn)品
免責(zé)聲明
- 凡本網(wǎng)注明“來(lái)源:化工儀器網(wǎng)”的所有作品,均為浙江興旺寶明通網(wǎng)絡(luò)有限公司-化工儀器網(wǎng)合法擁有版權(quán)或有權(quán)使用的作品,未經(jīng)本網(wǎng)授權(quán)不得轉(zhuǎn)載、摘編或利用其它方式使用上述作品。已經(jīng)本網(wǎng)授權(quán)使用作品的,應(yīng)在授權(quán)范圍內(nèi)使用,并注明“來(lái)源:化工儀器網(wǎng)”。違反上述聲明者,本網(wǎng)將追究其相關(guān)法律責(zé)任。
- 本網(wǎng)轉(zhuǎn)載并注明自其他來(lái)源(非化工儀器網(wǎng))的作品,目的在于傳遞更多信息,并不代表本網(wǎng)贊同其觀點(diǎn)和對(duì)其真實(shí)性負(fù)責(zé),不承擔(dān)此類作品侵權(quán)行為的直接責(zé)任及連帶責(zé)任。其他媒體、網(wǎng)站或個(gè)人從本網(wǎng)轉(zhuǎn)載時(shí),必須保留本網(wǎng)注明的作品第一來(lái)源,并自負(fù)版權(quán)等法律責(zé)任。
- 如涉及作品內(nèi)容、版權(quán)等問(wèn)題,請(qǐng)?jiān)谧髌钒l(fā)表之日起一周內(nèi)與本網(wǎng)聯(lián)系,否則視為放棄相關(guān)權(quán)利。